flow Schematic
Specifications
Model | ZYR-I-10L | ZYR-II-10L |
Feed Water Requirement | ||
Source | Tap water | Tap water |
TDS* | <200ppm | <200ppm |
Pressure | 0.1~0.5MPa | 0.1~0.5MPa |
Temperature | 5~45℃ | 5~45℃ |
Purification Water( TYPE II) | ||
Conductivity At 25℃ | < 2-10μS/cm | / |
Resistivity At 25℃ | / | 8-10MΩ.cm |
Bacteria Rejection | >99% | >99% |
Productivity Rate | 1.8L/M | 1.8L/M |
Heavy metalion **** | / | <0.1ppb |
Particulate(≥0.02um) | / | <1pc/ml |
Bacteria*** | / | <1cfu/ml |
Desalination rate | >99% | >99% |
Electrical Requirements | ||
Electrical Voltage | 220V-240 V | 220V-240V |
Electrical Frequency | 50HZ/60HZ | 50HZ/60HZ |
Dimension and weight | Length×Width×Height:530×400×590mm, 40Kg | |
Standard configuration | Main body (Including 1 set of cartridge)+15 liters pressure tank | |
Can be selected flow rate ( L/H) | 10L/H, 20L/H, 40L/H, 60L/H,100L/H |
Features and Advantages
○ ZYR.series is the sole leading brand of Chinese Good Instrument in lab water area. EspecialLY working with Biochemical analyzer
○ It is the most representative products in Chinese lab water market.
○ RO water, deionized water, or ultrapure water respectively.
○ Cartridges replacing alarm function, based on time and water quality, show cartridges´ used and residual life.
○ Multiple alarm function: no feed water, full water, water quality´s standard exceeding, and cartridge life ending.
○ Auto self-flushing of RO membrane function, extend RO membrane´s life.
○ System sanitizing procedure, achieve the disinfection of ultrapure water´s tube and valve.
○ External water tanks is optional to meet different need and assure ample water-supply.
○ Whole plastic shell with high-strength, avoid rusting and keep clean, to meet GLP standard.
○ 2 door and easy-to-replacing cartridge design, convenient to maintain system and replace cartridges.
○ Tube and adapter with NSF authorization and top quality, reduce TOC level and assure ultrapure water´s quality.
○ Optimized pretreatment (including PP fiber, KDF and active carbon cartridge), effectively protect RO membrane.
○ High quality RO module membrane, ensure long life, stable operation and high desalinization rate.
Typical Scientific Applications
○ Qualitative analysis
○ Animal watering
○ Aquariums
○ Autoclave feed
○ Glassware Washing/rinsing
○ biochemical analysis of chemical reagents;
○ buffer conventional physical and chemical detection;
○ Wahsing machines
○ biochemical analyzer