Purify urban tap water directly to
Laboratory grade pure water and ultrapure water;
Meet the international GB/T33087-2016
1.Full plastic ABS engineering plastic case, anti-acid corrosion;
2.RO water conductivity online monitoring, UP ultra-pure water resistivity online monitoring, water temperature online monitoring;
3.With quantitative water setting function (water tank storage);
4.With historical database query function;
5.With USB interface;
6.TOC online display (optional);
7.Pretreatment column / ultra purification column monitoring, alarm function (optional).
Technical Parameters:
Model | ZYMICRO-I-10L | ZYMICRO-II-10L | ZYMICRO-III-10L | ZYMICRO-IV-10L |
Output(25°C)* | 10 liters/hour | |||
Flow rate | Up to 1.8 liters/minute(with pressure tank) | |||
Feed water requirements | Tap water, temperature:5-45°C,pressure:1.0-4.0Kgf/cm2 | |||
RO water quality | ||||
Condutivity(25°C) | 2-10μs/cm | |||
Ion rejection rate | 99% (new RO membrane) | |||
Organic rejection rate | >99%, when MW>200 Dalton | |||
Particles and bacteria rejection rate | >99% | |||
Ultrapure water quality | ||||
Resistivity(25°C) | 18.2MΩ.cm | |||
Bacteria | <1cfu/ml | <1cfu/ml | <0.1cfu/m | <0.1cfu/ml |
Particle(>0.1μm) | <1pc/ml | <1pc/ml | <1pc/ml | <1pc/ml |
Heavy metal ion | <0.1ppb | <0.1ppb | <0.1ppb | <0.1ppb |
Endotoxin(Pyrogens) | N/A | N/A | N/A | ≤ 0.001EU/ml |
TOC Level*** | < 10 ppb | < 10 ppb | < 5 ppb | < 5 ppb |
Dimension and weight | Length×Width×Height:530×400×590mm, 40Kg | |||
Electrical requirements | AC110-240V, 50/60Hz | |||
Terminal disinfection filter | N/A | yes | yes | yes |
UV lamp | N/A | N/A | yes | yes |
UF | N/A | N/A | N/A | yes |
Standard configuration | Main body (Including 1 set of cartridge)+10 liters pressure tank | |||
Can be selected flow rate ( L/H) | 10L/H, 20L/H, 40L/H, 60L/H, 80L/H, 100L/H, 120L/H |